You can not select more than 25 topics
Topics must start with a letter or number, can include dashes ('-') and can be up to 35 characters long.
98 lines
2.6 KiB
98 lines
2.6 KiB
|
4 days ago
|
[Board]
|
||
|
|
PortName=COM5
|
||
|
|
BaudRate=115200
|
||
|
|
ReadTimeout=5000
|
||
|
|
PreConnectCommand=x00o
|
||
|
|
ConnectCommand=x00c_001101:owt28006727ea97c7801
|
||
|
|
ReadIdCommand=x00c_001101:ow2800326003e
|
||
|
|
PostReadIdCommand=x00o
|
||
|
|
CalDefaultCommand=x00c_001001:owt28006727ea97c7801
|
||
|
|
EndToken=<end>
|
||
|
|
DtrEnable=false
|
||
|
|
RtsEnable=false
|
||
|
|
|
||
|
|
[BarcodeScanner]
|
||
|
|
Enabled=true
|
||
|
|
PortName=Auto
|
||
|
|
BaudRate=9600
|
||
|
|
ReadTimeout=500
|
||
|
|
IdleCommitMilliseconds=250
|
||
|
|
DtrEnable=true
|
||
|
|
RtsEnable=true
|
||
|
|
|
||
|
|
[StartSignal]
|
||
|
|
; USB-6501
|
||
|
|
Enabled=true
|
||
|
|
Connection=Ni6501
|
||
|
|
PhysicalChannel=Dev1/port0/line0
|
||
|
|
ActiveState=Low
|
||
|
|
PollIntervalMilliseconds=50
|
||
|
|
; 0 = wait until the equipment start button is pressed, without a time limit.
|
||
|
|
TimeoutMilliseconds=0
|
||
|
|
PostSignalDelayMilliseconds=2000
|
||
|
|
RequireInactiveBeforeStart=true
|
||
|
|
|
||
|
|
[Equipment]
|
||
|
|
; V 출력: Keysight 34465A DMM 값
|
||
|
|
; A 출력: Keysight E36233A Power Supply CH1 Current 값
|
||
|
|
; USB 장비는 Keysight Connection Expert 또는 NI MAX에서 보이는 VISA 주소를 Resource
|
||
|
|
Timeout=5000
|
||
|
|
SettleMilliseconds=300
|
||
|
|
PreBoardCommand=VOLT 5, (@2);OUTP ON, (@2)
|
||
|
|
PreBoardDelayMilliseconds=2000
|
||
|
|
LogoutCommand=OUTP OFF, (@1)
|
||
|
|
|
||
|
|
; 34465A DMM
|
||
|
|
VoltageConnection=LAN
|
||
|
|
VoltageResource=
|
||
|
|
VoltageIdnMatch=34465A
|
||
|
|
VoltageHost=192.168.10.13
|
||
|
|
VoltagePort=5025
|
||
|
|
VoltageSetupCommand=
|
||
|
|
VoltageReadCommand=MEAS:VOLT:DC?
|
||
|
|
|
||
|
|
; Separate resistance meter. Change its address/commands here.
|
||
|
|
; Read after V/current and CH1 output OFF. Normal response: Ohm -> kOhm; overload: 999 + OL status.
|
||
|
|
; Resistance is required for inspection; Disabled stops the inspection with an error.
|
||
|
|
ResistanceConnection=LAN
|
||
|
|
ResistanceResource=
|
||
|
|
; Optional model identifier to validate with *IDN? (leave blank until the model is known).
|
||
|
|
ResistanceIdnMatch=
|
||
|
|
ResistanceHost=192.168.10.34
|
||
|
|
ResistancePort=5025
|
||
|
|
ResistancePortName=
|
||
|
|
ResistanceBaudRate=115200
|
||
|
|
ResistanceTerminator=LF
|
||
|
|
ResistanceReadCommand=MEAS:RES?
|
||
|
|
ResistanceSetupCommand=
|
||
|
|
; Query the CURRENT power supply to confirm CH1 output is OFF before resistance measurement.
|
||
|
|
ResistancePowerOffQuery=OUTP? (@1)
|
||
|
|
ResistanceSettleMilliseconds=300
|
||
|
|
|
||
|
|
; E36233A Power Supply
|
||
|
|
CurrentConnection=LAN
|
||
|
|
CurrentResource=
|
||
|
|
CurrentIdnMatch=E36233A
|
||
|
|
; Change this to the E36233A LAN IP if the instrument uses a different address.
|
||
|
|
CurrentHost=192.168.10.12
|
||
|
|
CurrentPort=5025
|
||
|
|
CurrentPortName=
|
||
|
|
CurrentBaudRate=9600
|
||
|
|
CurrentTerminator=LF
|
||
|
|
CurrentSetupCommand=OUTP ON, (@1)
|
||
|
|
CurrentCleanupCommand=OUTP OFF, (@1)
|
||
|
|
CurrentReadCommand=MEAS:CURR? (@1)
|
||
|
|
|
||
|
|
; COM 포트로 잡히는 장비 예시:
|
||
|
|
; VoltageConnection=Serial
|
||
|
|
; VoltagePortName=Auto
|
||
|
|
; VoltageBaudRate=115200
|
||
|
|
|
||
|
|
; LAN 장비 예시:
|
||
|
|
; VoltageConnection=Tcp
|
||
|
|
; VoltageHost=192.168.0.10
|
||
|
|
; VoltagePort=5025
|
||
|
|
|
||
|
|
; E36233A 파워서플라이에서 CH1 전압까지 읽고 싶을 때 예시:
|
||
|
|
; VoltageReadCommand=MEAS:VOLT? CH1
|